7 standard filters Built-in thermal printer, USB for external printer Back spectrophotometry
Performance:
1.Back spectro photometry, 7 hard membrane filters, separate signal
process circuit for each wavelength,
2.filters select completed electronically, no mechanical movement
Hermetic solid-state optical system, low contamination of dust and
dampness, long life time of hard membrane filters
3.Titanium alloy flow cell with small capacity and minimize
cross-contamination
4.200000 patients test data and 300 test programs can be saved
5.Adopt 7” touch screen,800*400 pixels
Main Test Items List
Liver function: ALT AST ALP T-BIL D-BIL TP
ALB TTT CHE Nh3 γ-GT
Renal function: BUN CRE UA CO2-CP
Lipids: T-CHO TG HDL-C LDL-C
Cardiac function: CK CK-MB LDH HDBH
Others: GLU K Na Cl Ca
TECHNICAL SPECIFICATIONS
Analysis method: kinetics, end point, two point
Light source: halogen lamp 12V/10W
Optical system: 340-800nm
7 standard filter: 340, 405, 492, 510, 546, 578,
630nm wavelength, accuracy:±2nm
Abs range: 0.000-4.000Abs, resolution: 0.001Abs
Flow cell: titanium quartz alloy, 32uL; optical path: 10mm
TEMPERATURE
PELTIER elements, 25℃,30℃,37℃, Room,±0.1℃
Precision: error<1%, drift<5mAbs/hr
Printer: thermal printer, 57mm width paper
Display: 7” touch screen,800×400 pixels
CPU: Samsung S5PV210, CortexTM-A8, 1GHz
Memory: DDR2 SDRAM, 256MB Flash: 4GB
Interface: 3 USB ports ,1 RS232 serial port
Dimensions: 400(L)x290(W)x165(H)mm
WORKING CONDITIONS
Temperature: 10℃-30℃
Humidity: 30%-80%
Power supply: AC100-240V,50/60Hz